Features:
Two measuring methods: magnetic induction (F) and eddy current (N). When F probe is equipped, it can measure the non-magnetic coating in magnetism substrate, when N probe is equipped, it measure the non-conductive coating on non-magnetism substrate
6 types of probes are available for various applications:F400,F1,F1/90,F10,N1,CN02
2 measurement modes: continuous / single
5 statistical ways: Mean values / Max. values / Min. values / testing numbers.(No.) / standard deviations (S.DeV)
640 measured values can be stored
Integrated with printer for convinient use
Low battery indication
2 switch off modes: manual and auto
Technical Specification:
Measuring range |
Refer to the table below |
Probes available |
|
Tolerance |
|
Minimum resolution: |
|
Measuring condition |
|
Operation language |
English |
Standards |
DIN, ISO, ASTM,BS |
Calibration |
Zero and foil calibration |
Statistics |
Number of measurements, mean, standard deviation, |
Data memory |
640 readings |
Limits |
Adjustable with alarm |
Interface |
RS-232 |
Working temperature |
0-40℃ |
power supply |
NiMHrechargeable batteries 1.25V |
Dimensions |
230mm×86mm×47mm |
Weight |
Approx. 530g |
|
|||||
|
Optional probes and technical specification:
Probe model |
F400 |
F1 |
F1/90º |
F10 |
N1 |
CN02 |
|||
Operating principle |
Magnetic induction |
Eddy current |
|||||||
Measuring range (µm) |
0-400 |
0-1250 |
0-10000 |
0 to 1250 μm |
10~200 |
||||
Low range resolution (µm) |
0.1 |
0.1 |
10 |
0.1 |
1 |
||||
Accuracy |
One-point calibration (µm) |
±(3%H+1) |
±(3%H+10) |
±(3%H+1.5) |
±(3%H+1) |
||||
Two-point calibration (µm) |
±[(1~3)H%+0.7] |
±[(1~3)H%+1] |
±[(1~3)%H+10] |
±[(1~3)%H+1.5] |
- |
||||
Measuring conditions |
Min curvature of the min area (mm) |
Convex |
1 |
1.5 |
Flatten |
10 |
3 |
Flatten |
|
Diameter of the min area (mm) |
φ3 |
φ7 |
φ7 |
φ40 |
φ5 |
φ7 |
|||
Critical thickness of substrate (mm) |
0.2 |
0.5 |
0.5 |
2 |
0.3 |
unlimited |
Your Review: Note: HTML is not translated!
Rating: Bad Good
Enter the code in the box below: